17.Sync Pattern Start Position

Screens

Measurement Sync

This field is only displayed when Burst Type is set to User Def. It selects the starting bit position of a user defined synchronization pattern.

Range

0 to 147.

18.Sync Status This field displays problems that were detected during digital demodulation or DSP analyzer measurements.

 

See Also

Making Measurements: Solving Problems Messages

19. Trig Delay

This field sets the time delay between a valid trigger event and the beginning of a

 

measurement.

 

 

Range

0 to 5000.00 μs.

20. Trig Qual

This field selects whether a trigger qualifier will be used.

 

Choices

Normal, no trigger qualifier is used.

 

 

RF Power, trigger events will only be valid when RF power is

 

 

detected at the selected input. If no RF Power is present at the time

 

 

of the trigger, the measurement will re-arm after seven timeslots.

 

 

NOTE

If no measurements are being made while in continuous (Cont), check Trig Qual (see

 

Screens:RF Generator/RF Analyzer). If it is set to RF Power, change it to Normal.

 

 

 

21.Trig Source This field selects the source of the trigger. When Activated (see Screens: Cell Configuration), triggers for digital demodulation are generated internally.

Choices

Ext Demod selects triggering through RP_DMOD_TRIG on the

 

rear-panel SYSTEM BUS CONNECTOR.

 

RF Rise selects triggering when the rising edge of an RF signal is

 

detected. RF Rise is used with pulsed RF input signals.

 

Ext Meas selects triggering through the MEASURE TRIGGER IN

 

front panel connector.