Making Measurements

Measurements

After making one type of measurement, another can be made by simply pressing CELL CNTL and, using the knob, selecting the next measurement of your choice.

Peak Carrier Power The peak transmitter carrier power averages the transmitter carrier power for a

Measurement

single burst. This average is calculated over the time that the data information bits

 

are transmitted.

 

Select the Pwr Zero field, disconnect external source and lower BCH power, (3),

NOTE

 

see Figure 2-6 on page 2-14, to zero the power meter before any signal is applied.

 

Method

 

 

The peak transmitter carrier power value is displayed on the cell control screen.

 

After you have set up a call (Active Cell) or manually aligned the Agilent

 

8922M/S to the mobile phone (Test Mode/CW Generator), the Peak Power

 

reading is displayed in the center of the screen (1).

 

Below the CALL STATUS area of the cell control screen are reports from the

 

mobile phone. These appear only when using Active Cell. They indicate the TX

 

Level which the phone is transmitting and the RX Level and RX Quality of the

 

received signal, (2), see Figure 2-6 on page 2-14.

2

1

3

Figure 2-6

Peak Power Measurements

2-14

Page 42
Image 42
Agilent Technologies S GSM, 8922M manual Method, Peak Power Measurements