Agilent Technologies 8922M, S GSM manual Configuring, Set Parity to none, Set Data Bits to

Models: 8922M S GSM

1 381
Download 381 pages 1024 b
Page 303
Image 303

 

 

Instrument BASIC

 

 

Configuration and Instrument Control

Configuring the

1

Access the Agilent 8922M/S’s I/O CONFIGURE screen.

Agilent 8922M/S

2

Set Serial In field to Inst to allow the Agilent 8922M/S’s IBASIC controller

 

 

 

to accept characters from a PC or ASCII terminal.

 

3

Set IBASIC Echo to On.

 

4

Set Inst Echo to On.

 

5

Set the Serial Baud to 4800. (Baud can be altered as required by your

 

 

terminal.)

 

6

Set Parity field to None.

 

7

Set Data Length to 8 bits.

 

8

Set Stop Length to 1 bit.

 

9

Set Rcv Pace to None.

 

10

Set Xmt Pace to Xon/Xoff.

Configuring Your

Configuring an ANSI Terminal

Terminal or PC

1

Select ANSI operating mode.

 

 

2

Set Baud Rate to 4800 (if this rate is not available on your terminal, set it to a

 

 

rate that can be selected on the Agilent 8922M/S’s I/O CONFIGURE screen.

 

3

Set Parity to none.

 

4

Set Data Bits to 8.

 

5

Set AnqAck to no (or none).

 

6

Set Receive/Transmit Pacing to match the Agilent 8922M/S’s settings.

 

Your terminal may have additional fields available for different configurations, but

 

should be able to communicate with the Agilent 8922M/S if these settings are made.

Configuring an IBM-Compatible PC With HP AdvanceLink

HP AdvanceLink is a popular PC terminal emulator used to emulate a variety of terminals. If you are using a different terminal emulator program on a PC, configure it using the above settings.

8-7

Page 303
Image 303
Agilent Technologies 8922M, S GSM manual Configuring, Set Parity to none, Set Data Bits to, Set AnqAck to no or none