Spectrum Analyzer, Main Controls (Option 006 Only)

2.

Center Freq

This field sets the center frequency when the RF Analyzer’s Hop Mode is set to

 

 

Non-Hop.

 

 

 

Range

10.0 to 1015.0 MHz.

3. Marker

The two fields display the frequency and the level at the current marker position.

 

 

See Also

Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG, Units

4. Max Hold

 

 

 

5.

Meas Reset

Selecting Meas Reset will erase any accumulated measurements used for calculating

 

 

a final result, and re-start the measurement process for the following functions:

 

 

HI LIMIT

 

 

 

LO LIMIT

 

 

 

AVG

 

 

 

See Also

Keys: HI LIMIT, LO LIMIT, AVG

6.

Ref Level

This field is the reference level entry field for the selected RF input port.

 

 

Range

RF IN/OUT: 18.9 dBm to +44.0 dBm. AUX RF IN: 55.0

 

 

 

 

dBm to +23.0 dBm.

7.

Span

This field sets the frequency span.

 

 

Range

5 kHz to 4 MHz (overrange to 10 MHz).

4-123