Instrument BASIC

Entering and Editing Programs

Other PROG Commands

Two additional PROG Commands are used to prepare the Agilent 8922M/S’s

IBASIC Controller RAM for receiving programs.

OUTPUT Addr;”PROG:DEL”

deletes any programs currently residing in RAM.

OUTPUT Addr;”PROG:DEF #0”

defines the address in RAM where a downloaded program will be stored.

Preparing the Agilent 8922M/S to Receive Programs

1Configure the GPIB port as described in this chapter, ‘GPIB Configuration For Programming” on page 8-5 .

2Access the IBASIC Controller screen.

Downloading A Program Into the Agilent 8922M/S

This procedure assumes your Agilent 8922M/S’s GPIB address is 14. If it is not, change the address in the following procedure to match your instrument’s address.

1Load the IBASIC program to be downloaded into your controller.

2Enter these commands to transfer the program to the Agilent 8922M/S:

10DIM LINE$[200], FILE_NAME$[120]

20Addr=714

30

INPUT ”NAME OF (ASCII) IBASIC FILE TO DOWN-

LOAD?”, FILE_NAME

$

 

 

40

ASSIGN @FILE TO FILE_NAME$; FORMAT ON

 

50

ON END @FILE GOTO DONE

 

60

OUTPUT Addr;”PROG:DEL”

 

70

OUTPUT Addr;”PROG:DEF #0”

 

80

WHILE (1)

 

90

ENTER @FILE; LINE$

 

100

OUTPUT Addr; LINE$

 

110

END WHILE

 

120

DONE: !

 

130

OUTPUT Addr; ” ” END

 

140

PRINT ”Done with down-load”

 

150

END

 

Figure 8-6

Download Program for Computers using GPIB

To verify that your code was downloaded, type in the command- OUTPUT 714;”PROG:EXEC ‘LIST ’”. Your program should be listed on the Agilent 8922M/S’s IBASIC Controller screen.

8-18

Page 314
Image 314