Making Measurements

 

Agilent Technologies 8922M/S Operating Modes

 

To configure the Agilent 8922M/S to the desired operating mode, carry out the

 

following instructions, referring to Figure 2-1 on page 4 for the position of the

 

fields.

 

• Ensure(1) is set to mobile phone type you require (GSM900, E-GSM, DCS1800,

 

or PCS1900).

 

 

NOTE

For use with DCS1800 or PCS1900, refer to the HP/Agilent 83220A/E Users Guide

 

which explains how to configure the Agilent 8922M/S to test other mobile phone

 

formats.

 

 

Use the knob to move the cursor to the lower field below the Operating Mode(2).

Push the knob and select the desired Operating Mode (ACTIVE CELL, TEST MODE, or CW GENERATOR) from the menu which appears at the bottom right-hand side of the screen.

1

2

Figure 2-1

Agilent 8922M/S Operating Modes

In addition to the three main operating modes, there are three other modes available. These are:

ACTIVE CELL+

TEST MODE+

CW GENERATOR+

These “+” modes have all the features of the other modes but include extra diagnostic information on the screen. This information is designed to help advanced users in fault-finding mobile phones.