Making Measurements

CW GENERATOR

CW GENERATOR

To enter this mode, select CW GENERATOR as described in the section titled “Agilent 8922M/S Operating Modes”. The CW Generator mode has the same measurement capabilities as the Test Mode but replaces the GSM BCH and forced TCH signals with a single unmodulated RF carrier. The frequency and amplitude of the Continuous Wave (CW) signal is controlled by the Channel, Amplitude, and the Frequency fields (1). When a GSM channel number is entered in the Channel field, this automatically updates the generator Frequency field below it. If it is necessary to generate a non GSM channel frequency, enter a value directly into the frequency field. This range varies from 10 MHz through 1000 MHz. A direct frequency entry over-rides the Channel field above. The unmodulated RF signal can be used for any general purpose application and can be particularly useful for checking some aspects of the mobile phone’s receiver where a CW signal is needed.

1

Figure 2-4

CW Generator

2-12

Page 40
Image 40
Agilent Technologies S GSM, 8922M manual CW Generator