Instrument BASIC

Loading, Storing, and Running

Figure 8-5

Making a Test Procedure File

Running a Test

1 Select the Run Test field.

Sequence

2 Follow directions and prompts on the Agilent 8922M/S screen according to the

 

 

test sequence being run.

 

3 When testing is complete, the Agilent 8922M/S will respond to front panel or

 

remote input. If at any time you need to stop testing, press the front-panel

 

 

 

key.

 

 

CANCEL

 

 

 

 

 

 

 

 

NOTE

A program takes up to 3 minutes to load into the Agilent 8922M/S and is loaded when

 

the Run Test field is first pressed.

 

Press the front-panel

 

 

 

keys to abort from an error condition or to

 

SHIFT

CANCEL

 

abort from the program. When you abort from loading the program, you’ll need to

 

clear Agilent 8922M/S RAM memory in order to correctly re-load the program at a

 

later time.

 

To clear Agilent 8922M/S RAM, select and run the ROM program COPY_PL from

 

the Procedure field in the TESTS menu. (Running COPY_PL deletes all

 

SAVE/RECALL registers. So instead you may want to load another program, run it,

 

and then re-load the original program that was aborted.

 

 

 

 

 

 

 

 

8-15

Page 311
Image 311
Agilent Technologies 8922M, S GSM manual Making a Test Procedure File, Running a Test, Sequence