Making Measurements

Measurements

Table 3

Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to

 

 

Modulation

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power

 

 

Offset from Carrier (kHZ)

 

 

Control

 

 

 

 

 

 

 

 

 

0

100

 

200

250

400

 

600 to

 

Level

 

 

 

 

 

 

 

 

 

 

1800

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 (43 dBm)

0 dB

0.5 dB

 

-27.5 dB

-32 dB

-51 dB

 

-70 dB

 

 

 

 

 

 

 

 

 

 

 

0 (39 dBm)

0

0.5

 

-27.5

-32

-51

 

-66

 

 

 

 

 

 

 

 

 

 

 

0 (37 dBm)

0

0.5

 

-27.5

-32

-51

 

-64

 

 

 

 

 

 

 

 

 

 

 

5

0

0.5

 

-27.5

-32

-51

 

-60

 

(33 dBm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 4

Adjusted Limits Based on 3-Pole Filter Output RF Spectrum Due to Switching

 

 

 

 

 

 

 

 

 

Power Level (dBm)

 

Offset from Carrier (kHZ)

 

 

 

 

 

 

 

 

 

 

400

 

600

1200

1800

 

 

 

 

 

 

 

 

 

 

43

-3 dB

 

-17 dB

-19 dB

-24 dB

 

 

 

 

 

 

 

 

 

 

41

-5

 

-17

-19

-24

 

 

 

 

 

 

 

 

 

 

39

-7

 

-17

-19

-24

 

 

 

 

 

 

 

 

 

 

37

-9

 

-17

-19

-24

 

 

 

 

 

 

 

 

 

 

35

-11

 

-17

-19

-24

 

 

 

 

 

 

 

 

 

 

33

-13

 

-17

-19

-24

 

 

 

 

 

 

 

 

 

 

31

-15

 

-19

-21

-26

 

 

 

 

 

 

 

 

 

 

29

-17

 

-21

-23

-28

 

 

 

 

 

 

 

 

 

 

27

-17

 

-22

-25

-30

 

 

 

 

 

 

 

 

 

 

25

-17

 

-22

-27

-32

 

 

 

 

 

 

 

 

 

 

23

-17

 

-22

-29

-34

 

 

 

 

 

 

 

 

 

 

21

-17

 

-22

-30

-36

 

 

 

 

 

 

 

 

 

2-23

Page 51
Image 51
Agilent Technologies 8922M, S GSM manual 32 dB 51 dB 70 dB 39 dBm 27.5 37 dBm ≤ 33 dBm