4. Aux BCCH

NOTE

NOTE

5. BA

6.BCC

7.CA

Screens

Cell Configuration - GSM 900

This field allows selection of data and clock outputs.

This is not featured in the Agilent 8922S.

Choices

Off causes the front-panel MODULATION IN/OUT DATA and

 

CLOCK connectors to be inputs.

 

Adjacent causes data and clock signals to be output on the front-

 

panel MODULATION IN/OUT DATA and CLOCK connectors.

 

These signals can be connected to a 0.3 GMSK generator

 

(HP/Agilent 8657A/B Option 022.) A mobile station should be

 

able to camp on to the Aux BCCH.

The BCC (Base Station Colour Code) is displayed in this field and has a number plus 1 modulo 4 relationship with Serv Cell BCC.

The data output includes a midamble, which will result in discrete sidebands. If random data is desired, use an HP/Agilent 8904A Option 001 or 002 to generate random data and clock signals to the 0.3 GMSK generator.

See Also

Connectors and Timing Diagrams: CLOCK (MODULATION)

 

Connectors and Timing Diagrams: IN/OUT DATA

 

(MODULATION)

This field is a 124-element Boolean array defining which ARFCNs are in the base station allocation. The serving cell ARFCN does not have to be set to a 1 in this table.

Choices

Done exits the choices menu.

 

Position moves the cursor to the desired position in the array.

 

0 means the ARFCN is not selected.

 

1 means the ARFCN is selected.

This field sets the base-station color-code portion of the BSIC (base station identity code) for the serving cell.

Range

0 through 7

This field is a 124-element Boolean array defining which ARFCNs are in the cell allocation table. It is used to define the ARFCNs used in the TCH hop sequence.

Choices

Done exits the choices menu.

 

Position moves the cursor to the desired position in the array.

 

0 means the ARFCN is not selected.

 

1 means the ARFCN is selected.

4-18

Page 124
Image 124
Agilent Technologies S GSM, 8922M manual Aux Bcch BCC, Range Through