Screens

 

 

CW Measurement

3. CW Power

This field displays the measured CW power. This measurement is valid only for

 

non-pulsed signals at the front-panel RF IN/OUT connector. This measurement is

 

made at the frequency entered in the Frequency field.

 

See Also

Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG,

 

 

METER, Units

4.Frequency (Expected Input)

5.Meas Reset (Meas Cntl)

This field sets the frequency of the signal to be measured. It needs to be within ±500 kHz to ensure the correct readings of amplitude (3) and frequency (2).

In the GSM band, the value that is entered should be set to the nearest 100 kHz.

Selecting Meas Reset will erase any accumulated measurements used for calculating a final result, and re-start the measurement process for the following functions:

HI LIMIT

LO LIMIT

AVG

This field resets the AF Analyzer measurements.

See Also

Keys: HIGH LIMIT, LO LIMIT, AVG

6.Pwr Zero (Meas This field zeros the power meter. RF power must be disconnected from the RF

Cntl)

IN/OUT port when executing this function.

7.RF Cnt Gate This field sets the frequency counter’s gate time.

 

Range

0.11 seconds to 1 second.

8. RF Input

This field selects the front-panel RF input for the RF analyzer.

 

Choices

RF IN/OUT

 

 

AUX RF IN