Agilent Technologies S GSM, 8922M manual Pulse Puls Rise Puls Fall

Models: 8922M S GSM

1 381
Download 381 pages 1024 b
Page 204
Image 204

 

Screens

 

 

Pwr Ramp, Top 2 dB

4. Mask

This field turns the amplitude mask on or off.

 

 

NOTE

In PCS 1900 mode only (requires HP/Agilent 83220A/E) this field has three choices;

 

Off, Narrow or Relax. Narrow refers to the old ETSI phase 1 power vs time mask.

 

Relax refers to the new ETSI phase 11 power vs time mask where the specifications

 

are relaxed at lower power levels.

 

 

5. Pk TX Pwr

This field displays the average power over the useful bits in the measured burst. It is

 

measured from the center of the first useful bit to the center of the last useful bit.

 

This measurement is only available when using the RF IN/OUT port.

 

See Also

Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG, Units

6. SyncStatus

This field displays any errors that occurred while trying to synchronize to the

 

demodulated data. During multi-burst measurement, this field indicates the

 

measurement progress. When the measurement is complete, the SyncStatus is

 

displayed.

 

 

See Also

Making Measurements: Solving Problems Messages

7. Trig Delay

This field sets the time delay between a valid trigger event and the beginning of a

 

measurement.

 

 

Range

0 to 5000.00 μs.

8. View

This field selects alternate views of the measurement.

 

Choices

Rise Edge

 

 

 

Top 2 dB

 

 

 

Fall Edge

 

 

 

Summary

 

If you have option 006 (spectrum analyzer) installed, you will also be offered the

 

following measurement screens:

 

Pulse

 

 

Puls Rise

 

 

Puls Fall

 

4-98

Page 204
Image 204
Agilent Technologies S GSM, 8922M manual Pulse Puls Rise Puls Fall