Making Measurements

If You Have Problems with a Measurement

Solution 3 - Level

Check the following fields on the RF Generator/RF Analyzer screen (press

SHIFT , CELL CNTL , (RFG/RFA) to gain access).

3Pulse field (a), (if signal is pulsed) Ext or Hop Trig should be selected.

4Amplitude field (b), for the expected amplitude

5Frequency field (c), for the correct frequency.

6RF Input field (d), for the correct connector choice.

7AGC Mode field (e). If either Open or Auto is the selected mode, check the value in the Open Loop DAC (f) Value field. If Closed is the selected mode, check that the burst is repetitive (at least one timeslot every two frames).

Check the actual amplitude of the input signal.

Perform a Peak Carrier Power measurement.