Installation Overview

4To guarantee the correct operation of the Agilent 8922M/S with all mobile phones, you should use a high-stability timebase.

This step details how to configure the Agilent 8922M/S for use with either the option 001, high-stability timebase, or an external reference signal.

A If you have option 001 installed and wish to use this as your reference, ensure

that the OPT 001 REF OUT field in the Configure screen is set to On. Access the Reference field in the Configure screen and select 10 MHz.

B If you wish to use an external reference signal, set the Reference field to 1, 2, 5, 10, or 13 MHz according to the frequency reference you intend to use.

A

B

Figure 1-4

High-stability timebase set-up

To customize your Agilent 8922M/S further, refer to Chapter 4, Screens.

1-7