Section 5: Printer Configuration

LCD DISPLAY

DEFINITION

 

 

 

GAP

[X.XXV]

GAP - When setting the “gap” threshold, the voltage shown on the top line of

INPUT

[X.XV]

the display must be measured with nothing but the backing in the sensor and

 

 

then again with a label still attached to the backing. The formula to be used for

 

 

setting the threshold is:

 

 

(High Voltage Level + Low Voltage Level) x 0.5 = Start Value

 

 

1. Insert a label still attached to the backing into the sensor and close the

 

 

Label Hold-Down. Record the voltage shown on the top line of the LCD

 

 

panel. This line should have the message “GAP” on the top line (DIP

 

 

switch DSW2-2 = OFF). Ensure the label is fully under the sensor.

 

 

2. Strip the label from the backing and insert the backing strip under the

 

 

sensor and close the Label Lid. Record the voltage shown on the top line

 

 

of the LCD panel. The voltage ranges measured should be within the

 

 

following ranges:

 

 

Backing with label = 2.0V to 3.5V

 

 

Backing without label = Less than 1.0V

 

 

If the measured values are outside this range, it may be difficult finding a value

 

 

that will work properly under all conditions. If this is the case, a higher quality

 

 

label may be needed to get adequate performance.

 

 

3. Calculate the starting point voltage using the formula.

 

 

4. Use the LINE and FEED keys to step the counter to the desired setting.

 

 

The reading will advance to a setting of 4.9 (the maximum voltage). If a

 

 

value of “0.0” is set, the printer will automatically set the level each time

 

 

the printer is powered on with labels loaded and the head is closed.

 

 

5. Once the setting is correct, pressing the ENTER key will accept the

 

 

setting and advance the next display.

M84Pro Service Manual

PN: 9001111A

Page 5-18

Page 51
Image 51
SATO M84PRO service manual Setting the threshold is