SATO M84PRO service manual Label Sensor Adjustment

Models: M84PRO

1 135
Download 135 pages 48.2 Kb
Page 108
Image 108

Section 8: Adjustment Procedures

NOTE: Refer to TP Test Module Usage in the Troubleshooting section for additional instruction on test module use.

4Turn dial (3) on test module (1) to position number 5.

5Connect the positive (+) probe of the multimeter to pin SIG (4) on test module (1).

6Connect the negative (-) probe of the multimeter to pin GND (5) on test module (1).

7Ensure the millimeter is set for DC voltage reading.

8Place the label backing only in position to be read by the sensor.

9Adjust VR4 potentiometer so that the multimeter value is less than 0.5 volts.

10Replace the label backing with a label in position to be read by the sensor.

11Regard the millimeter for a value of +1.0 volts more than the prior value read.

12Repeat steps 8 through 11 until the value is +1.0 volts.

13Test print labels to ensure proper function.

14Disconnect test module (1) from main circuit board (2) and replace housing covers.

NOTE: Figure 10-1 in the Diagrams & Schematics section provides guidance on housing cover installation.

84 +RS

BOARD

IEEE12

 

 

 

 

 

1A

 

 

 

 

2A

 

 

 

 

3A

 

 

0

 

4A

 

 

1

5A

 

8

 

 

 

 

 

 

 

6A

 

 

 

 

 

7

 

2

1B

 

 

 

 

3

6

 

 

2B

 

3

3B

 

 

 

 

5

4

 

4B

 

 

 

 

 

D SIG

 

5B

 

 

 

 

1

GN

 

6B

 

 

 

5 4 2

 

 

1A

 

 

2A

 

 

3A

0

 

4A

1

5A

8

 

 

6A

7

2

1B

 

 

 

 

2B

6

3

3B

5

4

4B

 

GND

SIG

5B

1

 

6B

 

 

Figure 8-9, Label Sensor Adjustment

M84Pro Service Manual

PN 9001111A

Page 8-12

Page 108
Image 108
SATO M84PRO service manual Label Sensor Adjustment