Performance Verification
Operating Mode Tests
The following procedures verify the operation of the Cont, Triggered and Gated modes.
NOTE. When you output signals from the CH1 or CH1 OUTPUT, check that the other OUTPUT ( CH1 or CH1 ) LED is off. If the other OUT-
PUT LED is on, push the CH1 or CH1 OUT button to turn off the output.
Check Cont Mode
Equipment | A 50 W SMA5coaxial cable,5a5SMA female6BNC male5adapter, and |
required | an oscilloscope5TDS784D. |
|
|
Prerequisites | The AWG710 Arbitrary Waveform Generator must meet the |
| prerequisites listed on |
|
|
Do the following steps to install the test hookup and set the test equipment controls:
1.Use a 50 W SMA coaxial cable to connect the AWG710 Arbitrary Waveform Generator CH1 output connector to the oscilloscope CH1 input connector (see Figure 4±5).
AWG710 Arbitrary Waveform Generator | Oscilloscope5TDS784D |
50 W SMA5coaxial cable
Figure
2.Set the oscilloscope controls as follows:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . | CH1 |
CH1 coupling . . . . . . . . . . . . . . . . | DC |
CH1 scale . . . . . . . . . . . . . . . . . . | 200 mV/div |
CH1 input impedance . . . . . . . . . . | 50 W |
Horizontal |
|
Sweep . . . . . . . . . . . . . . . . . . . . . | 100 ns/div |
Trigger |
|
SMAFemale6BNC Male adapter
AWG710 Service Manual |