Adjustment Procedures

Enabling Service Mode All the adjustment procedures are obtained through the Service menu which is a submenu of the Utility menu. To enter this menu, you must enable the Service mode using the switch on the A10 Connector board. Perform the following step:

Power on the waveform generator by pushing ON/STBY button while pressing the service switch on the A10 Connector board. See Figure 5±1.

If you perform the adjustment procedure without removing the cabinet, you must insert the adjustment tool (Table 5±4, Item 7) in the hole on the left side of the waveform generator to engage the service switch.

Toggle the ON/STBY button to STBY to disable the Service mode.

￿10 ￿onnector board

Service switch

￿djustment tool

Figure 5-1: Accessing the service switch

5-4

AWG710 Service Manual