Performance Verification

Figure 4-26: Initial waveform output

b.Close SW5 of the ground closure to generate an event signal on the EVENT IN connector strobe pin.

c.Verify that the oscilloscope displays the DC waveform as shown in Figure 4±27.

Figure 4-27: DC waveform output when the SW5 is closed

d.Open SW5 of the ground closure to degenerate the strobe signal on the EVENT IN connector strobe pin.

e.Verify that the DC waveform is displayed on the oscilloscope.

f.Close SW5 of the ground closure again.

4-46

AWG710 Service Manual