Adjustment Procedures

Table 5-2: Adjustments

Adjustments

Refer to

 

 

Calibration

Page 4-3

 

 

￿eference Clock Frequency

Page 5-6

 

 

Marker level

Page 5-7

 

 

Magic Frequency

Page 5-10

 

 

Before doing the adjustments, note the following requirements.

Personnel This procedure is only to be performed by trained service technicians.

Warm%Up Period This waveform generator requires a 20 minute warm-up time in a 20_ C to 30_ C environment before it is adjusted. Any adjustments done before the operating temperature has stabilized may cause performance errors.

Self Calibration Calibrate the waveform generator using the self calibration routine. Refer to the Performance Verification section for instructions.

Performance These adjustment procedures require you to load a file from the Performance Check/Adjustments Files Check/Adjustment disk supplied with the waveform generator. Refer to

Adjustment Instructions in this subsection for information about this file.

Test Equipment Table 5±3 lists all test equipment required to adjust the waveform generator.

5-2

AWG710 Service Manual