Preparation for Use

Check Voltage Settings Check that you have the proper electrical connections. The AWG710 Arbitrary Waveform Generator operates within the following power supply voltage and frequency ranges:

Line voltage range

100 - 240 V

 

 

Line frequency

48 - 63 Hz

 

 

Maximum power

220 VA

 

 

Connect Power Cord Connect the proper power cord from the rear panel power connector to the power system.

NOTE. The AWG710 Arbitrary Waveform Generator is shipped with a 115 V power cord. If the AWG710 Arbitrary Waveform Generator is to be used with 230 V power, the power cord must be replaced with one appropriate for the power source used. See Table 2±2 for the available power cord types.

Standby Power Push the PRINCIPAL POWER SWITCH (shown in Figure 2±1) on the rear panel of the instrument. Power is now applied to the instrument standby circuitry. Once the instrument is installed, leave the PRINCIPAL POWER SWITCH on and use the ON/STBY switch, located on the front panel, to toggle the instrument between ON and STBY.

PRINCIPAL POWER SWITCH

Fuse

Power connector

Figure 2-1: Rear panel power switch, fuse holder, and power connector

AWG710 Service Manual