Performance Verification

Trigger signal CH2

Waveformoutput CH1

Figure 4-16: Trigger signal and waveform output (+5 V check 1)

e.Change the pulse waveform high level to 5.35 V by using the arrow keys.

f.Check that the sine waves are displayed on the oscilloscope as shown in Figure 4±17.

Triggersignal CH2

Waveformoutput CH1

Figure 4-17: Trigger signal and waveform output (+5 V check 2)

7.Follow the substeps below to check the trigger level accuracy at -5 V:

a.Push -, 5 and ENTER keys in this order.

b.Set the DC waveform low level of the function generator to -5.35 V by using the arrow keys.

AWG710 Service Manual

4-39