Performance Verification
Event Input and Enhanced Mode Tests
These procedures check the event input signals and enhanced mode operation.
NOTE. The event input check with strobe off and the strobe input check are structured as a continuous test. After the Check Event Input with Strobe Off test is complete, the Check Strobe Input test uses the connections and oscilloscope settings from the Check Event Input with Strobe Off test.
Check Event Input with Strobe Off
Equipment | A650W SMA6coaxial cable,6a6SMA female7BNC male6adapter and |
required | oscilloscope6TDS784D, and custom7made ground closure. See Figure |
| |
|
|
Prerequisites | The AWG710 Arbitrary Waveform Generator must meet the |
| prerequisites listed on |
|
|
Do the following steps to install the test hookup and set the test equipment controls:
1.Use a 50 W SMA coaxial cable and a
AWG710 Arbitrary Waveform Generator | Oscilloscope6TDS784D |
50 W SMA6coaxial cable | SMAFemale7BNC |
| Male adapter |
Figure
2.Connect the ground closure to the EVENT IN connector on the AWG710 Arbitrary Waveform Generator rear panel.
3.Set the oscilloscope controls as follows:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . | CH1 |
CH1 coupling . . . . . . . . . . . . . . . . | DC |
CH1 scale . . . . . . . . . . . . . . . . . . | 200 mV/div |
CH1 input impedance . . . . . . . . . . | 50 W |
Horizontal |
|
Sweep . . . . . . . . . . . . . . . . . . . . . | 500 ns/div or 400 ns/div |
AWG710 Service Manual |