![](/images/backgrounds/275910/275910-267115x1.png)
Performance Verification
6.Follow the substeps below to set the AWG710 Arbitrary Waveform Generator controls and to select the waveform file:
a.Push UTILITY
b.Push SETUP
c.Load the MODE.WFM file.
Refer to Loading Files on page 4±10 for file loading procedures.
d.Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connectors are on.
e.Push the FORCE TRIGGER button.
Verify that the oscilloscope displays a
7.Follow the substeps below to check the triggered mode with external triggering:
a.Turn on the function generator output.
b.Verify that the oscilloscope displays a
Retain the test hookup.
Waveform CH1 output
Trigger CH2 signal
Figure
AWG710 Service Manual |