Operating Basics

Table 2-15: Setup output parameter operations (cont.)

Bottom

Side

 

button

button

Description

 

 

 

Horizontal

Clock

Sets the clock sample rate from 10 kS/s to 200 MS/s.

 

Clock Src

Sets the clock source to either Internal or External. When set to

 

 

Internal, 10 MHz internal reference clock is used. When set to

 

 

External, a valid external clock signal is DC to 200 MHz with a

 

 

voltage level of 0.4 Vpk6pk and up to ±2 V.

 

Clock Ref

Sets the reference clock source to either Internal or External. A

 

 

valid external clock signal is 10 MHz ±0.1 MHz with a voltage

 

 

level of 0.2 V to 3.0 Vpk6pk.

 

Skew...

Sets the skew between Ch1, Ch2 and CH3 within -2.52 ns to 2.52

 

 

ns with 70 ps steps.

 

 

 

Run Mode

Continuous

Displays the Run Mode side menu for setting the instrument run

 

Triggered

mode. Refer to the Run Mode Menu subsection on page 3-4 for

 

Gated

an explanation of the different run modes.

 

Enhanced

 

 

 

 

Trigger

Source

Sets trigger source to Internal or External. If External is selected,

 

 

only the interval item is selectable.

 

Slope

Sets the trigger slope or gate polarity to Positive or Negative.

 

Level

Sets the trigger signal level. The trigger level range is ±5.0 V in

 

 

0.1 V increments.

 

Impedance

Sets the external trigger input line impedance to either 50 W or

 

 

1￿kW.

 

Interval

Sets trigger interval from 1.0 ms to 10.0 s.

 

 

 

Save/Re6

Save Setup

Save the setup parameters set by SETUP window as a setup file.

store

Restore

Restore a setup file.

 

 

Setup

 

 

 

 

Outputting a Waveform To output a loaded waveform, push the CH (1) OUT and/or CH (1 )_OUT front panel button(s), and then the front panel RUN button. The LEDs near each button light to indicate they are enabled. The instrument outputs the waveform(s) depending on the Run mode. You can turn either or both channel outputs on or off while the instrument is running by pushing the CH (1) OUT or the CH (1) OUT buttons. To stop the waveform output, push the RUN button so that the LED light turns off.

AWG710 Service Manual

2-45