Performance Verification

AWG710 Arbitrary Waveform Generator

50 W BNC,coaxial cable

Input,connector +,BNC-to-dual banana adapter +,50 W BNC,coaxial,cable

DMM

CH1,output,connector +,SMA￿Ma￿-BNC￿Fe￿ adapter +,BNC 50 W,precision terminator +,50 W,BNC coaxial cable

Figure 4-9: Amplitude accuracy initial test hookup

2.Set the DMM controls as follows:

Mode . . . . . . . . . . . . . . . . . . . . . . . . . .

VDC

Range . . . . . . . . . . . . . . . . . . . . . . . . .

Auto

Input . . . . . . . . . . . . . . . . . . . . . . . . . .

Front

3.Follow the substeps below to set the AWG710 Arbitrary Waveform Generator controls and to select the sequence file:

a.Push UTILITY (front-panel)￿System(bottom)￿Factory Reset (side)￿OK (side).

b.Push SETUP (front-panel)￿Run Mode (bottom)￿Enhanced (side). The AWG710 Arbitrary Waveform Generator is set to enhanced mode.

c.Load the AMP1.SEQ file.

Refer to Loading Files on page 4±10 for file loading procedures.

4.Push the RUN and CH1 output buttons.

The LEDs above the RUN button and CH1 output connector are on.

5.Do the following substeps to set the AWG710 Arbitrary Waveform Generator amplitude and confirm the offset setting:

a. Push VERTICAL MENU (front-panel)￿Amplitude (side).

AWG710 Service Manual

4-23