Performance Verification

c.Open the SW2 of the ground closure to degenerate the event signal.

d.Verify that the oscilloscope displays the waveform shown in Figure 4±21.

9.Check the EVENT IN connector pin 2 input:

a.Close SW3 of the ground closure to generate an event signal on the EVENT IN connector pin 2.

b.Verify that the oscilloscope displays the waveform shown in Figure 4±24.

Figure 4-24: Waveform output when the SW3 is closed

c.Open SW3 of the ground closure to degenerate the event signal.

d.Verify that the oscilloscope displays the waveform in Figure 4±21.

10.Check the EVENT IN connector pin 3 input:

a.Close the SW4 of the ground closure to generate an event signal on the EVENT IN connector pin 3.

b.Verify that the oscilloscope displays the waveform shown in Figure 4±25.

4-44

AWG710 Service Manual