Performance Verification
c.Verify that the oscilloscope displays the waveform as shown in Figure 4±22 and that the waveform amplitude is the half size of Figure 4±21.
Figure
d.Open SW1 of the ground closure to degenerate the event signal.
e.Verify that the oscilloscope displays the waveform in Figure 4±21.
8.Check the EVENT IN connector pin 1 input:
a.Close SW2 of the ground closure to generate an event signal on the EVENT IN connector pin 1.
b.Verify that the oscilloscope displays the waveform as shown in Fig- ure 4±23.
Figure
AWG710 Service Manual |