![](/images/backgrounds/275910/275910-267144x1.png)
Performance Verification
1/4 Clock Frequency(and(10 MHz Reference Input Tests
These procedures check the 10 MHz reference input function of the AWG710
Arbitrary Waveform Generator .
Equipment | A 50<W SMA coaxial cable, A 50<W BNC<coaxial cable, a<SMAfe= |
required | male=BNCmale<adapter,<a frequency counter, and a function |
| generator. |
|
|
Prerequisites | The AWG710 Arbitrary Waveform Generator must meet the |
| prerequisites listed on |
|
|
Do the following steps to install the test hookup and set the test equipment controls:
1.Use a 50 W SMA coaxial cable and SMA
2.Use a 50 W BNC coaxial cable to connect the AWG710 Arbitrary Waveform Generator 10 MHz REF IN connector to the function generator output connector (see Figure 4±28).
AWG710 Arbitrary Waveform
Generator rear panel
Function Generator AFG310 | Frequency Counter |
1/4<CLOCK<OUT | 10<MHz<REF<IN |
|
50 W BNC<coaxial cable | 50 W SMA<coaxial cable | SMAFemale=BNC |
|
| Male adapter |
Figure
a.Set the frequency counter controls as follows:
INPUT A
Coupling . . . . . . . . . . . . . . . . . . . . AC
FUNCTION . . . . . . . . . . . . . . . . . . . . . . A FREQ
Gate time . . . . . . . . . . . . . . . . . . . . . . . 0.2 s
Trigger Level . . . . . . . . . . . . . . . . . . . . . 0 V
AWG710 Service Manual |