Performance Verification

Figure 4-1: Diagnostic menu

Push the Execute Diagnostic side button to execute all of the AWG710

Arbitrary Waveform Generator diagnostics automatically.

NOTE. Push the Abort Diagnostic button (side menu) to stop a selected diagnostic after it is completed. This button does not stop the diagnostic while it is processing. The Abort button stops any further diagnostics from operating.

The internal diagnostics do an extensive verification of AWG710 Arbitrary Waveform Generator functions. While this verification progresses, the screen displays the clock icon. When finished, the resulting status appears on the screen.

Verify that no failures are found and reported on-screen. If the diagnostics terminates without error, Pass is displayed instead of the - - -. If a value is displayed, meaning an error is detected, consult a qualified service technician for further assistance.

Push any bottom or menu button (other than UTILITY) to exit the diagnostic screen.

4-4

AWG710 Service Manual