Performance Verification

CH1 input impedance . . . . . . . . . .

50 W

CH2 input impedance . . . . . . . . . .

1 M W

Horizontal

 

Sweep . . . . . . . . . . . . . . . . . . . . .

1 ms/div

Trigger

 

Source . . . . . . . . . . . . . . . . . . . . .

CH1

Coupling . . . . . . . . . . . . . . . . . . . .

DC

Slope . . . . . . . . . . . . . . . . . . . . . .

Positive

Level . . . . . . . . . . . . . . . . . . . . . .

+100 mV

Mode . . . . . . . . . . . . . . . . . . . . . .

Auto

a.Set the function generator controls as follows:

Function . . . . . . . . . . . . . . . . . . . . . . . .

DC

Mode . . . . . . . . . . . . . . . . . . . . . . . . . .

Continuous

Output . . . . . . . . . . . . . . . . . . . . . . . . .

Off

b.Push UTILITY (front-panel)￿System(bottom)￿Factory Reset (side)￿OK (side).

c.Push SETUP (front-panel)￿Run Mode (bottom)￿Gated (side).

d.Load the TRIG.WFM file.

Refer to Loading Files on page 4±10 for file loading procedures.

4.Push the RUN and CH1 OUT buttons.

The LEDs above the RUN button and CH1 output connector are on.

5.Set the trigger level to 5 V by the following substeps:

a.Push SETUP (front-panel)￿Trigger(bottom)￿Level (side).

b.Push 5 and ENTER keys in this order.

6.Set the trigger input signal level to 4.65 V by the following substeps below:

a.Turn on the function generator output.

b.Select offset by pressing the OFFSET button of the function generator.

c.Set the DC waveform High level to 4.65 V by using the arrow keys.

d.Check that no waveform is displayed on the oscilloscope as shown in Figure 4±16.

4-38

AWG710 Service Manual