Performance Verification

b.Set the function generator (AFG310) controls:

Fun￿tion . . . . . . . . . . . . . . . . . . . . . . . .

Squ￿￿￿

Mo￿￿ . . . . . . . . . . . . . . . . . . . . . . . . . .

Continuous

P￿￿￿m￿t￿￿s

 

F￿￿qu￿n￿y . . . . . . . . . . . . . . . . . .

10 MHz

Amplitu￿￿ . . . . . . . . . . . . . . . . . . .

2.0 V into 50 W ￿4.0 V into 1 MW￿

O￿￿s￿t . . . . . . . . . . . . . . . . . . . . . .

0 V

Output . . . . . . . . . . . . . . . . . . . . . . . . .

On

Check￿1/4￿Clock frequency

1.Follow the substeps below to set the AWG710 Arbitrary Waveform Generator controls and select the waveform file:

a.Push UTILITY (front-panel)￿System(bottom)￿Factory Reset (side)￿OK (side).

b.Load the MODE.WFM file.

Refer to Loading Files on page 4±10 for file loading procedures.

c.Push HORIZONTAL MENU (front-panel)￿Clock(side).

d.Push 2, 0, 0 and M (SHIFT+7) keys in this order or turn the general pur- pose knob to set the internal clock frequency to 200 MHz.

2.Push the RUN button.

The LED above the RUN button is on.

3.Verify that the frequency counter reading is 50 MHz ± 50 Hz (using the internal reference clock).

4.Retain the test hookup.

Check￿10MHz￿Reference

Input￿

1.Push HORIZONTAL MENU (front-panel)￿Clock Ref (side) so that the AWG710 Arbitrary Waveform Generator reference clock is set to External.

2.Verify that the frequency counter reading is 50.0 MHz ±10 kHz (using the external reference clock).

3.Change the function generator controls as follows:

F￿￿qu￿n￿y . . . . . . . . . . . . . . . . . . . . . .

10.1 MHz

4.Check that the frequency counter reading is 50.5 MHz ±10 kHz.

5.Turn the function generator output off and disconnect the function generator and frequency counter.

AWG710 Service Manual

4-49