Performance Verification
Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
These procedures check the accuracy of the AWG710 Arbitrary Waveform
Generator direct waveform outputs; amplitude and offset.
Check Amplitude
and
DC Offset
Equipment | A 504W BNC4coaxial cable, a 504W4precision terminator, an SMA |
required | male5BNC female4adapter, BNC female5to5dual banana adapter, |
| and a digital multimeter DMM. |
|
|
Prerequisites | The AWG710 Arbitrary Waveform Generator must meet the |
| prerequisites listed on |
|
|
Do the following steps to install the test hookup and set the test equipment controls:
1.Use a 50 W BNC coaxial cable, an
AWG710 Arbitrary Waveform Generator
50 W BNC4coaxial cable
CH14output4connector
+4SMAMale5BNCFemale adapter +4BNC 50 W4precision terminator +450 W4BNC coaxial cable
Input4connector +4BNC5to5dual banana adapter +450 W BNC4coaxial4cable
DMM
Figure
2.Set the DMM controls as follows:
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . | VDC |
Range . . . . . . . . . . . . . . . . . . . . . . . . . | 2 V |
Input . . . . . . . . . . . . . . . . . . . . . . . . . . | Front |
3.Push UTILITY
AWG710 Service Manual |