Adjustment Procedures

This section provides the procedures necessary to manually adjust the AWG710

Arbitrary Waveform Generator.

Overview

Description This subsection contains the following topics:

￿Overview. Basic information about adjustment requirements

￿Before Adjustments. General information about adjusting the waveform generator and the Performance Check/Adjustment disk files.

￿Adjustment Instructions. Procedures for manually adjusting the waveform generator.

Purpose Use the Adjustment Instructions subsection to return the waveform generator to conformance with the performance specified in the Specifications section. This procedure is not required to verify the performance of the waveform generator. To verify the performance of the instrument, refer to the Performance Verification section.

Adjustment Interval Perform the adjustment procedures every 12 months.

Adjustment After Repair After the removal and replacement of a module due to electrical failure, perform the adjustment procedures listed in Table 5±1.

Table 5-1: Adjustments required

Removed and replaced module

Adjustments required

 

 

A50 AWG boar￿ an￿ Flas￿ ￿￿sk

Clo￿k R￿￿. DAC, Cal￿brat￿on an￿

 

 

Mark￿r l￿v￿l, Ma￿￿￿ Fr￿q.

 

 

 

A71

Output boar￿

Cal￿brat￿on

 

 

 

A77

Att￿nuator boar￿

Cal￿brat￿on

 

 

Ot￿￿r mo￿ul￿s

No n￿￿￿ to a￿just

 

 

 

Adjustments There are three adjustment procedures listed in Table 5±2. You can do Calibration and Marker level adjustments without removing the cabinet.

AWG710 Service Manual

5-1