Performance Verification

c.Check that no waveform is displayed on the oscilloscope as shown in Figure 4±18.

Waveform CH1 output

Trigger CH2 signal

Figure 4-18: Trigger signal and waveform output (-5 V check 1)

d.Change the pulse waveform low level of function generator to -4.65 V by using cursor keys.

e.Check that sine waves are displayed on the oscilloscope as shown in Figure 4±19.

Waveform CH1 output

Trigger CH2 signal

Figure 4-19: Trigger signal and waveform output (-5 V check 2)

8.Push the RUN button and the CH1 button to turn off the RUN LED.

9.Turn off the function generator output to disconnect the function generator.

4-40

AWG710 Service Manual