Performance Verification

Source . . . . . . . . . . . . . . . . . . . . . CH1

Coupling . . . . . . . . . . . . . . . . . . . . DC

Slope . . . . . . . . . . . . . . . . . . . . . . Positive

Level . . . . . . . . . . . . . . . . . . . . . . +100 mV

Mode . . . . . . . . . . . . . . . . . . . . . . Auto

Do the following steps to set the AWG710 Arbitrary Waveform Generator controls and to select the waveform file:

1.Push UTILITY (front-panel)￿System(bottom)￿Factory Reset (side)￿OK (side).

2.Load the MODE.WFM file.

Refer to Loading Files on page 4±10 for file loading procedures.

3.Push the RUN and CH1 output buttons.

The LEDs above the RUN button and CH1 output connectors are on.

4.Check that the amplitude of the sine wave displayed on the oscilloscope is five vertical divisions and that a waveform of approximately one cycle per 2.5 horizontal divisions is displayed.

Check Triggered Mode The following table lists the equipment and prerequisites required to check the Triggered mode.

Equipment

Two 505W BNC5coaxial cables, a5505W SMA5coaxial cable,

required

a5SMA￿Fe￿6BNC￿Ma￿5adapter,5a5BNC6T5￿male to 2 females￿ adapter , a

 

function generator, and an oscilloscope5￿TDS784D￿.

 

 

Prerequisites

The AWG710 Arbitrary Waveform Generator must meet the

 

prerequisites listed on page￿4-8.

 

 

Do the following steps to install the test hookup and set the test equipment controls:

1.Use a 505W BNC coaxial cable and a BNC-T adapter to connect the function generator output connector to the AWG710 Arbitrary Waveform Generator TRIG IN connector. Refer to Figure 4±6.

2.Connect a second 505W BNC coaxial cable to the BNC-T adapter. Connect the opposite end of the coaxial cable to the oscilloscope CH2 input.

3.Use a 50 W SMA coaxial cable and SMA (female)-BNC (male) adapter to connect the AWG710 Arbitrary Waveform Generator CH1 output connector to the oscilloscope CH1 input connector.

AWG710 Service Manual

4-17