Performance Verification
Source . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . +100 mV
Mode . . . . . . . . . . . . . . . . . . . . . . Auto
Do the following steps to set the AWG710 Arbitrary Waveform Generator controls and to select the waveform file:
1.Push UTILITY
2.Load the MODE.WFM file.
Refer to Loading Files on page 4±10 for file loading procedures.
3.Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connectors are on.
4.Check that the amplitude of the sine wave displayed on the oscilloscope is five vertical divisions and that a waveform of approximately one cycle per 2.5 horizontal divisions is displayed.
Check Triggered Mode The following table lists the equipment and prerequisites required to check the Triggered mode.
Equipment | Two 505W BNC5coaxial cables, a5505W SMA5coaxial cable, |
required | a5SMAFe6BNCMa5adapter,5a5BNC6T5male to 2 females adapter , a |
| function generator, and an oscilloscope5TDS784D. |
|
|
Prerequisites | The AWG710 Arbitrary Waveform Generator must meet the |
| prerequisites listed on |
|
|
Do the following steps to install the test hookup and set the test equipment controls:
1.Use a 505W BNC coaxial cable and a
2.Connect a second 505W BNC coaxial cable to the
3.Use a 50 W SMA coaxial cable and SMA
AWG710 Service Manual |