Adjustment Procedures
Magic Frequency This procedure uses the general purpose knob on the front panel to set the Magic Frequency parameter.
Equipment | One oscilloscope Item 1 |
Required | One SMA 50 W coaxial cable Item 4 |
| |
| SMA Fe to BNC Ma 50 W adapter Item 5 |
|
|
Do the following steps to connect the test equipment and set test equipment controls:
1.Use a SMA coaxial cable and a SMA (female) to BNC (male) adapter to connect the waveform generator CH1 Output connector to the CH1 input connector on the oscilloscope.
AWG710 Arbitrary Waveform Generator | Oscilloscope TDS784D |
CH1 output | SMA 50 W coaxial cable |
|
Figure
2.Set the Oscilloscope controls as follows:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . | CH 1 |
CH 1 coupling . . . . . . . . . . . . . . . . | DC |
CH 1 scale . . . . . . . . . . . . . . . . . . | 200 mV/div |
CH 1 input impedance . . . . . . . . . . | 50 W |
Horizontal |
|
Sweep . . . . . . . . . . . . . . . . . . . . . | 200 ns/div |
Trigger |
|
Source . . . . . . . . . . . . . . . . . . . . . | CH 1 |
Coupling . . . . . . . . . . . . . . . . . . . . | DC |
Slope . . . . . . . . . . . . . . . . . . . . . . | Positive |
Level . . . . . . . . . . . . . . . . . . . . . . | 0 V |
Mode . . . . . . . . . . . . . . . . . . . . . . | Auto |
CH1 input
SMAFe4BNCMa adapter
AWG710 Service Manual |