Performance Verification

7.Do the following substeps to change the connection to check the MARKER1:

a.Disconnect the 50 W precision terminator from the MARKER1 OUT con- nector.

b.Disconnect the cable with SMA (male)±BNC (female) adapter and BNC 50 W terminator) from MARKER 1 OUT connector.

c.Connect the 50 W precision terminator to the MARKER 1 OUT (removed in step a).

d.Connect the adapter, terminator, and cable (removed in step b)

to the MARKER1 OUT connector.

8.Press FORCE EVENT where needed.

9.Repeat steps 5 and 6 for the AWG710 Arbitrary Waveform Generator

MARKER1 OUT.

10.Repeat step 5 through 8 for MARKER 2 OUT and MARKER2 OUT.

11.Disconnect the DMM.

4-52

AWG710 Service Manual