![](/images/backgrounds/275910/275910-267148x1.png)
Performance Verification
7.Do the following substeps to change the connection to check the MARKER1:
a.Disconnect the 50 W precision terminator from the MARKER1 OUT con- nector.
b.Disconnect the cable with SMA (male)±BNC (female) adapter and BNC 50 W terminator) from MARKER 1 OUT connector.
c.Connect the 50 W precision terminator to the MARKER 1 OUT (removed in step a).
d.Connect the adapter, terminator, and cable (removed in step b)
to the MARKER1 OUT connector.
8.Press FORCE EVENT where needed.
9.Repeat steps 5 and 6 for the AWG710 Arbitrary Waveform Generator
MARKER1 OUT.
10.Repeat step 5 through 8 for MARKER 2 OUT and MARKER2 OUT.
11.Disconnect the DMM.
AWG710 Service Manual |