![](/images/backgrounds/275910/275910-267147x1.png)
Performance Verification
2.Set the DMM controls as follows:
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . | VDC |
Range . . . . . . . . . . . . . . . . . . . . . . . . . | Auto |
Input . . . . . . . . . . . . . . . . . . . . . . . . . . | Front |
3.Follow the substeps below to set the AWG710 Arbitrary Waveform Generator controls and to select the sequence file:
a.Push UTILITY
b.Push SETUP
c.Load the AMP1.SEQ file.
Refer to Loading Files on page 4±10 for file loading procedures.
4.Push the RUN button.
The LEDs above the RUN button is on.
5.Do the following substeps to set the AWG710 Arbitrary Waveform Generator marker output high level setting:
a.Push VERTICAL MENU
b.Push 3, ., 0 and ENTER keys in this order or turn the general pur- pose knob to set the marker level to 3.0 V.
c.Verify that the marker 1 high level reading on the DMM screen falls within 2.75 V through 3.25 V.
d.Push the FORCE EVENT button.
6.Do the following substeps to set the AWG710 Arbitrary Waveform Generator marker output low level setting:
a.Push Marker 1 Low Level side button.
b.Push ±, 1, ., 1 and ENTER keys in this order or turn the general pur- pose knob to set the marker level to ±1.1 V.
c.Verify that the marker 1 Low level reading on the DMM screen falls within ±1.255 V through ±0.945 V.
d.Push the FORCE EVENT button.
AWG710 Service Manual |