Performance Verification

b.Disconnect the adapter, terminator and cable from the CH1 output con- nector.

c.Connect the adapter, terminator, and cable (removed in step b) to the CH1 output connector.

d.Push the CH1 output button to turn on the CH1 LED.

12.Repeat step 8 through step 10 for the AWG710 Arbitrary Waveform Generator CH1.

13.Push the CH1 output button to turn off the CH1 LED.

14.Push the RUN button to turn off the RUN LED.

15.Disconnect the DMM.

Check Pulse Rise Time

Equipment

A 501W SMA1coaxial cable and an oscilloscope ￿TDS820￿.

required

 

 

 

Prerequisites

The AWG710 Arbitrary Waveform Generator must meet the

 

prerequisites listed on page￿4-8.

 

 

Do the following steps to install the test hookup and set the test equipment controls:

1.Use the 50 W SMA coaxial cable to connect the AWG710 Arbitrary Waveform Generator CH1 output connector to the oscilloscope CH1 input connector (see Figure 4±11).

AWG710 Arbitrary Waveform Generator

Oscilloscope ￿TDS820￿

50 W SMA1coaxial cable

Figure 4-11: Direct DA output pulse rise time initial test hookup

2.Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . .

CH1

CH1 coupling . . . . . . . . . . . . . . . .

DC if applicable

CH1 scale . . . . . . . . . . . . . . . . . .

100 mV/div

Horizontal

 

Sweep . . . . . . . . . . . . . . . . . . . . .

500 ps/div

AWG710 Service Manual

4-29