Performance Verification

e.Push 1, 0, 0 and M (SHIFT+7) keys in this order or turn the general purpose knob to set the internal clock frequency to 100 MHz.

f.Push the RUN and CH1 output buttons.

The LEDs above the RUN button and CH1 output connector are on.

g.Push the FORCE TRIGGER button.

Verify that the oscilloscope displays sine waves while the FORCE TRIGGER button is pushed and that the output stops when the Force Trigger button is released.

4.Follow the substeps below to check the gated mode with the gate signal:

a.Set the oscilloscope trigger source to CH2 and change the trigger level to 1 V.

b.Turn on the function generator output.

c.Verify that the oscilloscope displays sine waves while the function generator gate signal amplitude is High level. See Figure 4±8.

Waveform output CH1

Gate CH2 signal

Figure 4-8: Relationship between gate signal and waveform output

d.Push SETUP (front-panel)￿Trigger(bottom)￿Negative (side).

This changes the AWG710 Arbitrary Waveform Generator trigger polarity to negative.

5.Verify that the oscilloscope displays sine waves while the function generator gate signal amplitude is Low level.

6.Turn off the function generator output and disconnect from the oscilloscope.

AWG710 Service Manual

4-21