Adjustment Procedures
9.Follow the substeps below to adjust the Marker 2 level:
a.Press Marker 2 Adj. Level
b.Turn the general purpose knob to adjust the waveform of marker low level voltage to 0.00 V within"50 mV on the oscilloscope display.
10.Press Save to save the setting.
11.Disconnect the oscilloscope.
AWG710 Service Manual |