Specifications

Table 1-11: Auxiliary inputs

Characteristics

Description

 

 

 

PV reference page

 

 

 

 

 

 

 

 

Trigger input 5

 

 

 

 

 

 

Connector

Rear panel BNC connector

 

 

 

 

 

 

 

 

 

 

Impedance

1 kW or 50 W

 

 

 

 

 

 

 

 

 

 

Polarity (Trigger mode)/Slope

POS (positive) or NEG (negative)

 

 

(Gated mode)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Input voltage range

"

 

W

load

 

 

 

 

10 V, into a 1 k

 

 

 

 

 

"5 V, into a 50 W load

 

 

 

 

 

 

 

 

 

 

Threshold

 

 

 

 

 

 

 

Level

-5.0 V to 5.0 V

 

 

 

 

 

 

 

 

 

 

 

 

 

nAccuracy

"

"

0.1 V

Page 4-37

 

 

 

5% of level

 

 

 

 

Resolution

0.1 V

 

 

 

 

 

 

 

 

 

Triggered mode

See Figure 1-1.

 

 

 

 

Minimum pulse width

10 ns, 0.2 V amplitude

 

 

 

 

 

 

 

 

 

Trigger holdoff time

x109.5 clocks + 500 ns

 

 

 

 

 

 

 

 

 

 

Delay to analog out, Typical

211.5 clocks + 17 ns (Output: Norm, Filter: Through)

 

 

 

 

 

 

 

Gated mode

 

 

 

 

 

 

Minimum pulse width

1152 clocks + 10 ns, 0.2 V amplitude

 

 

 

 

 

 

 

 

 

Gate holdoff time

x1920 clock + 20 ns

 

 

 

 

 

 

 

 

 

 

Delay to analog out, Typical

(1335 to 1499.5) clocks + 9 ns (Output: Norm, Filter: Through)

 

5The characteristics are specified at the end of the BNC cable (012+0482+00).

AWG710 Service Manual

1-9