Performance Verification
Pulse Response Tests (Normal Out), (except option 02)
This procedure checks the pulse response characteristics of the AWG710
Arbitrary Waveform Generator output waveforms at amplitudes of 1 V.
Equipment | A 503W SMA3coaxial cable and an oscilloscope TDS820. |
required |
|
|
|
Prerequisites | The AWG710 Arbitrary Waveform Generator must meet the |
| prerequisites listed on |
|
|
Do the following steps to install the test hookup and set the test equipment controls:
1.Use the 50 W SMA coaxial cable to connect the AWG710 Arbitrary Waveform Generator CH1 output connector to the oscilloscope CH1 input connector (see Figure 4±14).
AWG710 Arbitrary Waveform Generator | Oscilloscope TDS820 |
50 W SMA3coaxial cable
Figure
2.Set the oscilloscope controls as follows:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . | CH1 |
CH1 coupling . . . . . . . . . . . . . . . . | DC if applicable |
CH1 scale . . . . . . . . . . . . . . . . . . | 200 mV/div |
Horizontal |
|
Sweep . . . . . . . . . . . . . . . . . . . . . | 500 ps/div |
Trigger |
|
Source . . . . . . . . . . . . . . . . . . . . . | CH1 |
Slope . . . . . . . . . . . . . . . . . . . . . . | Positive |
Level . . . . . . . . . . . . . . . . . . . . . . | 0 V |
Mode . . . . . . . . . . . . . . . . . . . . . . | Auto |
3.Push UTILITY
AWG710 Service Manual |