MediSense PCx manual Data Upload

Models: PCx

1 151
Download 151 pages 28.1 Kb
Page 120
Image 120

Data Upload

9

 

 

Once the test results have been collected during a prescribed period of time, they will then be uploaded into the MediSense QC Manager Data Management System. The QC Manager Data Management System provides a simple and automated way to collect, report and transfer data to a personal computer (PC).

Data upload is initiated by the computer and requires no special steps by the operator, as long as the monitor is connected properly and the QC Manager Program is running. For the data upload to begin, either the monitor must be in the optional docking station or the communications cable must be plugged into the data port on the monitor. Once the connection is established, communication between the monitor and the PC will begin automatically.

The monitor can send and receive data. During data upload, the monitor cannot be used for testing. When this occurs, the monitor will display the following message, “Please Wait for Data Upload”. Testing is preempted while the monitor is sending or receiving data.

The technical details of the data upload are described in the QC Manager manual. Data is transferred in ASTM E-1381 format.

If the monitor is removed before data transmission is complete, no data will have been uploaded. The data will be uploaded as part of the next data transmission.

If a problem occurs with the data upload, an error message may appear on the monitor. Error messages are shown in the Troubleshooting section of this manual. Further troubleshooting information for any data upload question is provided in the QC Manager manual.

Error Messages:

9-1