Change Setup:

If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.

Use these steps to access the Change Setup menu:

What You See

What You Do

Comments

on the Display

 

 

 

 

 

1.Press On/Off to turn on the monitor.

The Abbott logo screen will appear for a few seconds followed by the product name, the software version screen and then the Test Mode menu.

2. Press Menu.

The Menu Mode menu

 

will appear.

3. Press 5 for Change

The display prompts the

Setup.

operator to enter a

 

password of up to 8

 

characters. The password

 

is displayed as asterisks

 

on the monitor.

4.Key in the password, then press Enter. If the password is entered incorrectly, the monitor displays error message: “Incorrect Password.” Press Clear twice, verify the password then re-enter the correct password and press

Enter.

The factory set password is 123. Once in the Change Setup Mode, it is suggested that the administrator change the password. This can be done in the System Setup/Set Password option.

8-5

Questions? Refer to Page 10-13 For Customer Care Information.

Page 65
Image 65
MediSense PCx manual Change Setup, On the monitor