What You See | What You Do | Comments | |
on the Display | |||
|
| ||
|
|
|
13.The operator can select one of the following options:
| Press 1 | – Next Patient. |
| Press 2 | – Repeat Test. |
|
|
|
(5.8 mmol/L) | Press 3 | – Patient History. |
|
|
|
If 1 is selected, the Scan Patient ID screen will appear. Return to step 4 to continue with testing on the next patient.
If 2 is selected, the Scan or Enter Strip Lot screen will appear. Return to step 5 to repeat the test.
If 3 is selected, the monitor will display the last test result for this patient as well as allow the operator to view previous results.
Refer to the section entitled Data Review for Patient History for further information.
Important: If the blood glucose result appears to be inconsistent (lower or higher than expected), there may be a problem with the test strip. Repeat the test using a new test strip. Results that are incorrect may have serious medical consequences. Consult your healthcare professional before making any changes to diabetes medication plans if:
•The blood glucose results are not consistent with the physical symptoms AND you have ruled out common errors in technique.