Control Test Procedure

Use the following procedure to perform control tests.

Note: The word “Strip” on the display always refers to the test strip. If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.

What You See

What You Do

Comments

on the Display

 

 

 

 

 

1.Press On/Off to turn on the monitor.

The Abbott logo screen will appear for a few seconds followed by the product name, software version and then the Test Mode menu.

2.Press 2 – Control Test.

3.Press Scan to scan the Operator ID barcode, or manually enter the Operator ID via the keypad, then press

Enter.

4.Scan or manually enter the low control solution lot number via the keypad, then press

Enter.

After the control test is selected, the Operator ID screen may appear, if the setting has been enabled.

If the Operator ID is not required, just press enter.

After the Operator ID is entered, the Low Level Solution Lot screen will appear.

The Strip Lot screen will appear next.

If the Unexpected Level screen appears, the operator may either:

1.Enter 1 to ReEnter the expected level.

2.Enter 2 to continue to test the unexpected level.

4-2

Page 24
Image 24
MediSense PCx manual Control Test Procedure