Data Review for Proficiency Data

The monitor stores the last 60 proficiency test results.

Please refer to the beginning of the Data Review for the step-by-step instructions to access the Data Review menu.

Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.

If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.

What You See

What You Do

Comments

on the Display

 

 

 

 

 

1.Press 3 – Proficiency Data.

2.The operator can select one of the following options:

Press 1 – Previous, view the previous test result.

(2.5 mmol/L) Press 2 – Next, view the next test result.

3.When finished reviewing the data, the operator can:

Press Clear to return to the Data Review screen.

Press Menu to return to the Menu Mode menu.

Press On/Off to turn off the monitor.

The monitor now shows the most recent proficiency test result.

The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.

The display shows the following:

s: Sample Type or ID o : Operator ID

0 : Comment Code

5-25

Questions? Refer to Page 10-13 For Customer Care Information.

Page 52
Image 52
MediSense PCx manual Data Review for Proficiency Data