MediSense PCx manual During Any Test, Reason may also be Too Long

Models: PCx

1 151
Download 151 pages 28.1 Kb
Page 132
Image 132

During Any Test:

Symptom/Error

Explanation

How To Respond

Message

 

 

 

 

 

The reason may also be: Too Long

The barcode just scanned was not read correctly. Possible reasons are:

• The red scanner beam did not pass across all the bars of the code.

• The scanner was not held between 3 and 12 inches from the barcode.

• The barcode was damaged or poorly printed.

• The barcode was not of the type specified for this ID or lot number.

The ID or lot number just scanned or entered is too long or too short, according to format defined during the setup of this monitor.

The ID or lot number just scanned or entered* is not on the list of acceptable IDs or lot numbers defined for this monitor.

1.Rescan the barcode or manually enter the ID or lot number.

2.Notify the system administrator. The problem may be the printing of the barcode or the setup.

1.Verify and re-enter the ID or lot number using the barcode scanner.

2.Manually enter the ID or lot number.

3.If the problem persists, notify the system administrator.

1.Verify and re-enter the ID or lot number using the barcode scanner.

2.Manually enter* the ID or lot number.

3.If the problem persists, notify the system administrator.

* Manual entry of strip lots with alpha-numeric barcodes is not supported.

10-7

Questions? Refer to Page 10-13 For Customer Care Information.